Thermo-fisher-scientific CyberScan pH 1500 User Manual

Browse online or download User Manual for Tools Thermo-fisher-scientific CyberScan pH 1500. Thermo Fisher Scientific CyberScan pH 1500 User Manual

  • Download
  • Add to my manuals
  • Print

Summary of Contents

Page 1 - CyberScan pH 1500

Instruction Manual CyberScan pH 1500 Bench pH/mV Meter Tech nologyMadeEasy... 68X292327 Rev 0 12/02

Page 2 - PREFACE

Instruction Manual CyberScan pH1500 8 3.2 Temperature Probe For more accurate pH measurement, a temperature probe (supplied with the meter) is us

Page 3 - TABLE OF CONTENTS

Instruction Manual CyberScan pH1500 9 3.3 Display / Keys Overview of the meter screen display and function key layout. Press std key to ini

Page 4 - INTRODUCTION

Instruction Manual CyberScan pH1500 10 3.4 Screen Display Familiarise yourself with the layout of the digital screen display.

Page 5 - GETTING STARTED

Instruction Manual CyberScan pH1500 11 stdsetupprintentersaveview 4 SETUP MENUS 4.1 Using the Meter The setup button brings you to the setup

Page 6 - 2.2 CONNECTORS

Instruction Manual CyberScan pH1500 12 pHViewpHView% SLOPEBUFFERselectpHselectSTABLEPrintPrintPrintPrintDataDataclearBUFFERP1.0View pH electrode o

Page 7

Instruction Manual CyberScan pH1500 13 4.1.1 P1.0 View pH Electrode Offset This setup menu allows you to view the pH electrode offset in mV afte

Page 8 - USING THE METER

Instruction Manual CyberScan pH1500 14 4.1.2 P2.0 View pH Electrode Slope This setup menu allows you to view the pH electrode slope in % after

Page 9

Instruction Manual CyberScan pH1500 15 4.1.3 P3.0 pH Buffer Group This setup option allows you to select from 3 different buffer groups, each

Page 10 - 3.2 Temperature Probe

Instruction Manual CyberScan pH1500 16 printenter 3. Press enter to make a selection.. 1 3 6 8 10 132 4 7 10 12

Page 11 - 3.3 Display / Keys

Instruction Manual CyberScan pH1500 17 4.1.4 P4.0 pH Resolution This setup option allows you to select various pH resolution either 0.1, 0.01 o

Page 12 - 3.4 Screen Display

PREFACE Thank you for selecting the Eutech Instruments CyberScan pH 1500 bench meter. The instruction manual serves to explain the use of the Cyb

Page 13 - SETUP MENUS

Instruction Manual CyberScan pH1500 18 stdprintenter3. Press enter key to make selection and use the ▲ or ▼ key to choose either 0.1 (one decimal

Page 14 - stored earlier

Instruction Manual CyberScan pH1500 19 4.1.5 P5.0 Temperature Unit This setup option allows you to select unit of measure for Temperature eithe

Page 15 - ELECTRODE

Instruction Manual CyberScan pH1500 20 printenterstd3. Press enter key to make selection and use the ▲ or ▼ key to choose either °C or °F. sel

Page 16

Instruction Manual CyberScan pH1500 21 printenterstd4.1.6 P6.0 Stability Indicator This setup option allows you to set the stability indicator t

Page 17

Instruction Manual CyberScan pH1500 22 4.1.7 P7.0 Baud Rate This setup option allows you to set the baud rate (bits per second) of the communi

Page 18 - 1 3 6 8 10 13

Instruction Manual CyberScan pH1500 23 3. Press enter key to make selection and use the ▲ or ▼ key to toggle between 4800, 9600, 19200 and 38400.

Page 19

Instruction Manual CyberScan pH1500 24 4.1.8 P8.0 Parity Bit This setup option allows you to set the parity bit of the communication protocol i

Page 20

Instruction Manual CyberScan pH1500 25 3. Press enter key to make selection and use the ▲ or ▼ key to toggle between 0 (none), 1 (odd) or 2 (even)

Page 21

Instruction Manual CyberScan pH1500 26 4.1.9 P9.0 Stop Bit This setup option allows you to set the stop bit of the communication protocol inter

Page 22

Instruction Manual CyberScan pH1500 27 3. Press enter key to make selection and use the ▲ or ▼ key to toggle between 1 or 2. saveviewsaveview

Page 23

TABLE OF CONTENTS 1 INTRODUCTION 2 1.1 Introducing the CyberScan Series 2 2 GETTING STARTED 3 2.1 Meter Overview 3 2.2 CONNECTORS 4 3 USING THE

Page 24

Instruction Manual CyberScan pH1500 28 4.1.10 P10.0 Print Data This setup option allows you to print current displayed data or data stored in th

Page 25

Instruction Manual CyberScan pH1500 29 4.1.11 P11.0 Clear Stored Data This setup option allows you to clear all stored data sets (from previou

Page 26

Instruction Manual CyberScan pH1500 30 stdprintenter3. Press enter key to make selection and use the ▲ or ▼ key to toggle between NO or YES. Selec

Page 27

Instruction Manual CyberScan pH1500 31 stdprintenter4.1.12 P12.0 Clear Buffer Values This setup option allows you to clear the standardized buff

Page 28

Instruction Manual CyberScan pH1500 32 View  press to set valuepress to acceptenterELECTRODEmV4.1.13 P13.0 View mV offset This setup opt

Page 29

Instruction Manual CyberScan pH1500 33 4.1.14 P14.0 Clear mV offset This setup option allows you to clear the standardized mV offset values. To

Page 30

Instruction Manual CyberScan pH1500 34 5 PH OPERATION 5.1 Standardization Because electrodes vary in their response, you must standardize or c

Page 31

Instruction Manual CyberScan pH1500 35 ) This meter allows for automatic calibration using up to 6 buffers. Press mode key to tog

Page 32

Instruction Manual CyberScan pH1500 36 A. Using Standard Buffer Group a. Press std key to access the standardize mode while you are in the pH

Page 33

Instruction Manual CyberScan pH1500 37 f. Repeat the above step c through e for the second and subsequent calibration points. Meter automatically

Page 34

Instruction Manual CyberScan pH1500 2 1 INTRODUCTION 1.1 Introducing the CyberScan Series Thank you for selecting the Eutech Instruments Cybe

Page 35

Instruction Manual CyberScan pH1500 38 stdB. Using Custom Buffer Values a. Press std to access the standardize mode while you are in the pH Measu

Page 36 - PH OPERATION

Instruction Manual CyberScan pH1500 39 5.2 pH Electrode Diagnosis Indicator There is an electrode diagnosis icon that indicates the condition

Page 37

Instruction Manual CyberScan pH1500 40 modesave6 mV OR RELATIVE mV OPERATION 6.1 Absolute mV Millivolts (mV) measurement is used primarily fo

Page 38

Instruction Manual CyberScan pH1500 41 modestd6.2 Relative mV The Relative mV mode helps you to standardize certain analytical and monitoring a

Page 39

Instruction Manual CyberScan pH1500 42 6.3 Temperature Calibration 1. Press mode to go to the absolute mV measure screen 2. Press std. this wil

Page 40

Instruction Manual CyberScan pH1500 43 7 MEASUREMENT 1. Immerse the electrode (along with temperature probe if available for the measurement of

Page 41

Instruction Manual CyberScan pH1500 44 saveprintentersaveview 8 MEMORY 8.1 Store Value Into Memory In any Measure mode, press ▲ save key t

Page 42 - 6.1 Absolute mV

Instruction Manual CyberScan pH1500 45 stdprintenter9 PRINT DATA 9.1 Printing Data Depending on the print option in the meter setup, pressing

Page 43 - 6.2 Relative mV

Instruction Manual CyberScan pH1500 46 10 pH THEORY The measurement of pH plays an important role in quantifying and controlling acidity and alk

Page 44 - 6.3 Temperature Calibration

Instruction Manual CyberScan pH1500 47 10.1 Measuring pH To measure pH, the meter receives a millivolts signal from a glass bulb electrode that

Page 45 - MEASUREMENT

Instruction Manual CyberScan pH1500 3 2 GETTING STARTED 2.1 Meter Overview

Page 46

Instruction Manual CyberScan pH1500 48 11 CLEANING This meter requires no regular maintenance, but it is recommended to occasionally wipe down th

Page 47 - PRINT DATA

Instruction Manual CyberScan pH1500 49 13 METER SPECIFICATIONS Display Custom LCD Screen Size 7.2 x 10.6 cm Measurement display height

Page 48

Instruction Manual CyberScan pH1500 50 14 ACCESSORIES Consult your Authorized Distributors for these items and other range of specialized pH ele

Page 49 - 10.1 Measuring pH

Instruction Manual CyberScan pH1500 51 Calibration Solutions EC-BU-1BT pH 1.68 calibration buffer (480ml bottle) EC-BU-4BT pH 4.01 calibration b

Page 50 - TROUBLESHOOTING

Instruction Manual CyberScan pH1500 52 15 WARRANTY Eutech Instruments supplies this bench meter with a 3-year warranty and 6-month warranty for e

Page 51 - METER SPECIFICATIONS

Instruction Manual CyberScan pH1500 53 16 NOTICE OF COMPLIANCE Warning This meter generates, uses, and can radiate radio frequency energy. If no

Page 52 - ACCESSORIES

For more information on Eutech Instruments products, contact your nearest Eutech Instruments distributor or visit our

Page 53 - Calibration Solutions

Instruction Manual CyberScan pH1500 4 2.2 CONNECTORS 1. Review the layout and arrangement of the rear connector panel.

Page 54 - WARRANTY

Instruction Manual CyberScan pH1500 5 2. Connect the electrode arm to the base. 3. Connect the power cable to the connector cable to the rea

Page 55 - NOTICE OF COMPLIANCE

Instruction Manual CyberScan pH1500 6 3 USING THE METER 3.1 Electrodes This meter allows you to use two types of electrodes: the conventional g

Page 56

Instruction Manual CyberScan pH1500 7 ) Do not discard the BNC shorting cap. 3. Rinse and blot-dry (don’t wipe) electrodes between each measur