Page 1 - INSTRUCTION MANUAL

EUTECH INSTRUMENTS PTE LTD. INSTRUCTION MANUAL LEAD ION ELECTRODE

Page 2 - TABLE OF CONTENTS

Low Level Lead Determination (using a pH/mV meter) This procedure is recommended for solutions with lead concentrations of less than 1.0X10-6M. If t

Page 3

Instruction Manual Lead Electrode 9reading has stabilized, record the mV reading and determine the concentration from the low level calibration cu

Page 4 - LEAD ION ELECTRODE

Titration Titration is a very accurate determination of total lead or sulfate ion concentration. This method makes use of the electrode as an endpoin

Page 5 - Electrode Preparation

Instruction Manual Lead Electrode 11Titration of Sulfate Titration of sulfate ion with lead perchlorate make use of the lead ion electrode as a s

Page 6 - Sample Requirements

6. The sulfate ion concentration from the unknown is calculated as follows: Vt Mt

Page 7 - Direct Measurement

Instruction Manual Lead Electrode 13ELECTRODE CHARACTERISTICS Reproducibility Electrode measurements reproducible to ±2% can be obtained if the

Page 8

30 30.07 40 31.07 50 32.06 If changes in temperature occur, the electrodes should be re-calibrated. The temperature range for the E

Page 9

Instruction Manual Lead Electrode 15 The lower limit of detection is influenced by the slight water solubility of the electrode pellet. Refer to F

Page 10

The activity, X, represents the effective concentration of free lead ion in the solution. Both bound, Cb, and free, Cf, lead ions are included in the

Page 11

Instruction Manual Lead Electrode 17TROUBLESHOOTING GUIDE The goal of troubleshooting is the isolation of a problem through checking each of the

Page 12

TABLE OF CONTENTS GENERAL INSTRUCTIONS ...1 In

Page 13 - Titration of Sulfate

Technique Be sure that the electrode's limit of detection has not been exceeded. Be sure that the analysis method is clearly understood and is co

Page 14

Instruction Manual Lead Electrode 19TROUBLESHOOTING HINTS Symptom Possible Causes Next Step Out of Range de

Page 15

Low Slope or standards contaminated prepare fresh standards No Slope or incorrectly made stan

Page 16 - Limits of Detection

Instruction Manual Lead Electrode 21SPECIFICATIONS Concentration Range: 10-1M to 10-6M Pb+2 (20,700 to 0.2 ppm Pb+2) p

Page 17 - Electrode Operation

Instruction Manual Lead Electrode 1GENERAL INSTRUCTIONS ...Error! Bookmark not defined. I

Page 18

EUTECH INSTRUMENTS PTE LTD. LEAD ION ELECTRODE INSTRUCTION MANUAL GENERAL INSTRUCTIONS Introduction Eutech Instruments Lead Ion Electrode is used

Page 19

Instruction Manual Lead Electrode 3Pb(ClO4)2·3H2O. Swirl the flask gently to dissolve the solid. Fill the flask to the mark with distilled water,

Page 20 - Technique

3. Adjust the meter to the concentration of the standard and fix the value in the memory according to the meter manufacturer's instructions. 4.

Page 21

Instruction Manual Lead Electrode 5The temperature of the standard solutions and of the sample solutions should be the same and below 80oC. About

Page 22

5. Using the semi-logarithmic graph paper, plot the mV reading (linear axis) against the concentration (log axis). Extrapolate the calibration curve

Page 23

Instruction Manual Lead Electrode 7Direct Measurement of Lead (using an ion meter) 1. By serial dilution of the 0.1M or l,000 ppm lead standard,