Instruction Manual
CyberScan DO 5000
Bench Dissolved Oxygen Meter
(Also Applicable for CyberScan PCD 5500)
Technolo
g
y
M
ade
E
as
y
.
.
.
68X292330
Rev. 2 07/03
Page view 0
1 2 3 4 5 6 ... 104 105

Summary of Contents

Page 1 - CyberScan DO 5000

Instruction Manual CyberScan DO 5000 Bench Dissolved Oxygen Meter (Also Applicable for CyberScan PCD 5500)

Page 2 - PREFACE

Instruction Manual CyberScan DO5000 7 Standby screen Touch anywhere to resumeE U T E C H INSTRUMENTS11:11 am

Page 3 - TABLE OF CONTENTS

Instruction Manual CyberScan DO5000 97 Feature Details DO 5000 PCD 5500 pH mode Range -2.000 to ±20.000 ا ~  Resolution 0.1/ 0.01/ 0.00

Page 4 - 1 INTRODUCTION

Instruction Manual CyberScan DO5000 98 Feature Details DO 5000 PCD 5500 General Password Protection ~  Input/ output DIN/ RS232  

Page 5

Instruction Manual CyberScan DO5000 99 12 ACCESSORIES Consult your Authorized Distributors for these items and other range of specialized pH e

Page 6 - 2 UNPACKING THE METER

Instruction Manual CyberScan DO5000 100 13 WARRANTY Eutech Instruments supplies the bench meters with a 3-year warranty and 6-month warranty f

Page 8 - 3.2 Dissolved Oxygen Probe

For more information on Eutech Instruments products, contact your nearest Eutech Instruments distributor or visit

Page 9

Instruction Manual CyberScan DO5000 8Meter Main Screen Select from the options to the rightDOBODOURSOURsystemE U T E C H INSTRUMENTSCyberSc

Page 10 - 11:11 am

Instruction Manual CyberScan DO5000 93.3 Touch Screen Operation The CyberScan DO 5000 meter operates with a state of art touch screen. The

Page 11 - E U T E C H

Instruction Manual CyberScan DO5000 10savemodeenterclosemoreback3.4 Button functions The touch screen of your CyberScan research DO meter h

Page 12 - 3.3 Touch Screen Operation

Instruction Manual CyberScan DO5000 11prevnextprintmeassetupeditclearstdTBSBSdeleteThis button accesses the standardization screen from the v

Page 13 - 3.4 Button functions

Instruction Manual CyberScan DO5000 124 SYSTEM SETUP SYSTEM SETUP OPTIONS- Set - Set - Set - Set - Set - Set - Display - Reset to DateBeep

Page 14

Instruction Manual CyberScan DO5000 13SYSTEM SETUP OPTIONS- Set - Set - Set - - Set - Set - Set - Display - Reset to DateBeeper StatusBarcode

Page 15 - System Setup

Instruction Manual CyberScan DO5000 14 Access system setup System setup can be accessed from the Setup screen (Page 12). The system setup fu

Page 16

Instruction Manual CyberScan DO5000 154.1 Set Date 1-15-02Set DateCurrent MM/DD/YYclearTouch to delete current dateUse to input

Page 17

Instruction Manual CyberScan DO5000 16This screen can be used to set the present date which will be displayed on the measure screens. This da

Page 18 - 4.1 Set Date

PREFACE Thank you for selecting the Eutech Instruments CyberScan DO 5000 bench meter. This meter measures dissolved oxygen and temperature (°C/

Page 19

Instruction Manual CyberScan DO5000 174.2 Set Time Set TimeCurrent 12HR:clearTouch to delete current timeUse to input the new ti

Page 20 - 4.2 Set Time

Instruction Manual CyberScan DO5000 18This screen can be used to set the present time which will be displayed on the measure screens. This ti

Page 21 - Current 24HR

Instruction Manual CyberScan DO5000 194.3 Set Beeper Status Current BEEPER STATUS- After STABLE Reading - After KEY Touch

Page 22 - 4.3 Set Beeper Status

Instruction Manual CyberScan DO5000 20 This screen allows you to turn on or off the beeper. You may choose to have an audible signal when the

Page 23

Instruction Manual CyberScan DO5000 21 4.4 Set Print Configuration Current PRINT CONFIGURATION- Baud Rate - Numb

Page 24 - 4.4 Set Print Configuration

Instruction Manual CyberScan DO5000 22You can adjust the print configuration of this meter from this screen. The configuration of the followi

Page 25

Instruction Manual CyberScan DO5000 234.4.1 Set Baud Rate STenterSet Baud RateCurrent BAUD RATE11060012002400480096001920038400300300 T

Page 26 - Baud Rate

Instruction Manual CyberScan DO5000 244.4.2 Set Number of Bits enterCurrent NUMBER OF BITSSet Number of Bits78 To Set Number of Bits

Page 27 - Number of Bits

Instruction Manual CyberScan DO5000 254.4.3 Set Stop Bits enterCurrent STOP BITSSet Stop Bits12 To Set Stop Bits 1. Access the Set S

Page 28 - Stop Bits

Instruction Manual CyberScan DO5000 264.4.4 Set Parity enterCurrent PARITYSet ParityNONEEVENODD To Set Parity 1. Access the Set Parit

Page 29 - Current PARITY

TABLE OF CONTENTS 1 INTRODUCTION 1 1.1 Introducing the CyberScan Series 1 2 UNPACKING THE METER 3 3 GETTING STARTED 4 3.1 Connectors 4 3.2

Page 30 - Current MODE

Instruction Manual CyberScan DO5000 274.4.5 Set Serial Output Mode Current MODESetSerial Output ModeenterTEXTDATA To Set Serial Outpu

Page 31 - Barcode Configuration

Instruction Manual CyberScan DO5000 284.5 Set Barcode Configuration Current MODESet Barcode ConfigurationenterOFFON Use this option, as

Page 32 - 4.6 Set Operator

Instruction Manual CyberScan DO5000 294.6 Set Operator Set OperatorCurrent OperatorA B C 1

Page 33

Instruction Manual CyberScan DO5000 30This option allows you to identify the user of the meter. This information can be saved in the meter’s

Page 34 - 4.7 Set Display Contrast

Instruction Manual CyberScan DO5000 31 4.7 Set Display Contrast Set Display ContrastTouch or to adjust contrast and then touch t

Page 35

Instruction Manual CyberScan DO5000 32This option allows you to change the contrast on the screen to improve the readability of the informati

Page 36

Instruction Manual CyberScan DO5000 334.8 Display Meter Information Unit Serial Number: XX999XXXSoftware Revision:

Page 37 - Reset to Factory Defaults

Instruction Manual CyberScan DO5000 344.9 Reset to Factory Defaults You are about to reset all parameters of the meter to factory defaul

Page 38 - D.O. Setup

Instruction Manual CyberScan DO5000 355 D.O. / BOD/ OUR/ SOUR SETUP D.O. SETUP OPTIONS- Set - Select - Select

Page 39 - 5.1 To Access Setup

Instruction Manual CyberScan DO5000 365.1 To Access Setup From the DO, BOD, OUR and SOUR screens 1. Touch DO, BOD, OUR, or SOUR from the m

Page 40 - 5.2 Set Sample ID#

Instruction Manual CyberScan DO5000 1 1 INTRODUCTION 1.1 Introducing the CyberScan Series Thank you for selecting a Eutech Instruments Cyb

Page 41

Instruction Manual CyberScan DO5000 375.2 Set Sample ID# Manual ID#enterclearA B C 1 D

Page 42 - D.O. Sample ID#

Instruction Manual CyberScan DO5000 38When this option is active, each time you touch print on the Measure Screen the DO values along with da

Page 43

Instruction Manual CyberScan DO5000 39 Set D.O. Sample ID#Sequential ID#clearTouch to delete current IDUse to input the new st

Page 44 - D.O. Calibration Mode

Instruction Manual CyberScan DO5000 40 Sequential ID# Assignment 1. Access the Set Sample ID# screen from the Setup screen. 2. Touch seq f

Page 45 - D.O. Auto Read Mode

Instruction Manual CyberScan DO5000 415.3 Select Calibration Mode This mode permits a choice between automatic and manu

Page 46

Instruction Manual CyberScan DO5000 425.4 Select Auto Read Mode (Only available in D.O. mode) Current MODESelect D.O. Auto Read Modeent

Page 47

Instruction Manual CyberScan DO5000 435.5 Set Configurations for BOD/ OUR and SOUR (Not applicable in D.O. mode) Set BOD Configuration

Page 48 - D.O. Stability Criteria

Instruction Manual CyberScan DO5000 44Set SOUR Configuration This option allows the user select parameters that define the criteria for the

Page 49 - D.O. Temperature Units

Instruction Manual CyberScan DO5000 455.6 Set Stability Criteria Current MODESelect D.O. Stability CriteriaAUTOSTenterFastSlowMedium

Page 50 - D.O. Measurement Units

Instruction Manual CyberScan DO5000 465.7 Set Temperature Units Current UNITSSelect D.O. Temperature UnitsenterCFK 1. Access the Selec

Page 51 - D.O Salinity

Instruction Manual CyberScan DO5000 2

Page 52 - Barometer

Instruction Manual CyberScan DO5000 475.8 Set Measurement Units Current UNITSSelect D.O. Measurement Units%satSTentermg/Lmbar%sat

Page 53

Instruction Manual CyberScan DO5000 485.9 Set the Salinity Value of Your Sample 1. Access the Set Salinity screen from the DO/ BOD/ OUR/ S

Page 54 - 5.11 Set Alarm Limits

Instruction Manual CyberScan DO5000 495.10 Set Barometer Set BarometerCurrent PressureclearTouch to delete current pressureUse to

Page 55

Instruction Manual CyberScan DO5000 50The DO 5000 has an internal barometer. When it is no longer reflects the true barometer pressure, it mu

Page 56 - 5.12 Set Print Criteria

Instruction Manual CyberScan DO5000 515.11 Set Alarm Limits Current LIMITS- D.O. Alarm OFF1500- D.O. Maximum

Page 57

Instruction Manual CyberScan DO5000 52This option allows you to set alarm limits for the DO/ BOD/ OUR/ SOUR measuring mode. If the DO value o

Page 58 - 5.13 Set Print Interval

Instruction Manual CyberScan DO5000 535.12 Set Print Criteria Current PRINT CRITERIA- Date/Time/Channel - Sample ID

Page 59

Instruction Manual CyberScan DO5000 54This screen allows you to select which criteria are printed with the measurement when you print the dat

Page 60 - D.O. Data Storage Criteria

Instruction Manual CyberScan DO5000 555.13 Set Print Interval Set D.O. Print IntervalCurrent INTERVALtimedUse to input the desired p

Page 61

Instruction Manual CyberScan DO5000 562. Touch MAN to set the meter for manual printing. Touch enter to accept the print interval mode and re

Page 62 - D.O. Display Resolution

Instruction Manual CyberScan DO5000 3 2 UNPACKING THE METER The following is a listing of what you should have received with your new Cyber

Page 63

Instruction Manual CyberScan DO5000 575.14 Set Data Storage Criteria Current DATA STORAGE CRITERIA- Date/Time - Sampl

Page 64 - D.O. Display Configuration

Instruction Manual CyberScan DO5000 58This screen allows you to select what criteria are stored in the meter’s memory with the measurement wh

Page 65

Instruction Manual CyberScan DO5000 595.15 Set Display Resolution Set D.O. Display ResolutionCurrent RESOLUTIONSelect desired display

Page 66 - STORED DATA

Instruction Manual CyberScan DO5000 60Here you can set the display resolution on the screen. Choose high for two decimal points in mg/L, and

Page 67

Instruction Manual CyberScan DO5000 615.16 Set Display Configuration Current DISPLAY CONFIGURATION- Last Standardization

Page 68

Instruction Manual CyberScan DO5000 62This particular function will allow you to choose what information you would like to be displayed on th

Page 69

Instruction Manual CyberScan DO5000 635.17 View Stored Data View STORED DATAData POINTSclear17enterSTSample IDOperatorDateUse to

Page 70 - 6 D.O. OPERATION

Instruction Manual CyberScan DO5000 64The meter has memory capacity of up to 250 data points. The view stored data screen allows you to sort

Page 71 - 6.1 DO Standardization

Instruction Manual CyberScan DO5000 65To sort by Date 1. Access the Date sort option from the View Stored Data screen. 2. Touch clear to

Page 72 - 8.15 mg/L

Instruction Manual CyberScan DO5000 66If an operator ID is entered and no data points are stored with that operator ID, you will see a messag

Page 73

Instruction Manual CyberScan DO5000 43 GETTING STARTED 3.1 Connectors 1. Review the layout and arrangement of the rear connector panel.

Page 74

Instruction Manual CyberScan DO5000 676 D.O. OPERATION MeasuresetupTouch to measure sampleormeasTouch to access standardize modestd 

Page 75

Instruction Manual CyberScan DO5000 68 6.1 DO Standardization The CyberScan DO 5000 has two standardization modes. There is auto- standardi

Page 76 - Touch to measure sample

Instruction Manual CyberScan DO5000 69StandardizesetupTouch to delete previous standardInsert electrode into standard and stirclear Touch

Page 77

Instruction Manual CyberScan DO5000 70To Standardize the meter in the Automatic Standardization Mode 1. Insert the DO probe into a BOD bott

Page 78

Instruction Manual CyberScan DO5000 71MeasuresetupTouch to measure sampleormeasTouch to access standardize modestd stdmeasmg/L8.21STABLE

Page 79 - Measure - BOD

Instruction Manual CyberScan DO5000 72To Standardize the meter in the Manual Standardization Mode 1. Insert the DO into your choice of stan

Page 80

Instruction Manual CyberScan DO5000 73 MeasuresetupTouch to measure sampleormeasTouch to access standardize modestd stdmeasmg/L8.2

Page 81 - Measure-BOD

Instruction Manual CyberScan DO5000 74Measuring DO with Auto Read ON The measure screen provides a readout of the current sample measurement

Page 82

Instruction Manual CyberScan DO5000 75 Select from the options to the rightDOBODOURSOURsystemE U T E C H INSTRUMENTSMeasuresetupTouch to

Page 83 - Description

Instruction Manual CyberScan DO5000 76 6.2 Biological Oxygen Demand (BOD) Measurement Measure - BODsetupTouch to begin measurementorstar

Page 84 - Bottle volume

Instruction Manual CyberScan DO5000 5 3.2 Dissolved Oxygen Probe This meter is equipped to use the BOD probe. 1. Carefully review the in

Page 85

Instruction Manual CyberScan DO5000 77With the CyberScan DO 5000, BOD calculations are automatic. When the BOD button is touched, a variety o

Page 86

Instruction Manual CyberScan DO5000 78To Measure the BOD of the Seed Initial DO Measurement 1. Access the BOD mode from the main mode scree

Page 87

Instruction Manual CyberScan DO5000 795. If the sample is new, the BOD first reading screen appears. Touch SEED. Measure-BODsetupTou ch

Page 88 - Predilution (1 to X)

Instruction Manual CyberScan DO5000 806. Key in a sample description using the alphanumeric touch pad, SEED, for example. 7. Touch enter to

Page 89 - BOD Results

Instruction Manual CyberScan DO5000 81 8. Key in the volume of the SEED sample using the alphanumeric touch pad. Touch enter to accept the S

Page 90

Instruction Manual CyberScan DO5000 82Final DO Measurement 1. Access the BOD mode from the main mode screen 2. Once the DO reading is sta

Page 91

Instruction Manual CyberScan DO5000 83To Measure BOD of a Sample Initial DO measurement 1. Access the BOD mode from the main mode screen

Page 92

Instruction Manual CyberScan DO5000 84 5. If the sample is new, the BOD first reading screen appears. Touch SAMPLE. Measure-BODsetupTou ch

Page 93

Instruction Manual CyberScan DO5000 85STABLENot StandardizedPredilution (1 to X)1 2 3 4 5

Page 94

Instruction Manual CyberScan DO5000 86BOD ResultsSeed ID: 1, SEEDAPredilution:Seed Volume:Bottle Volume:Initial DO:Final DO:Delta DO:print8

Page 95 - 8 TROUBLE SHOOTING

Instruction Manual CyberScan DO5000 6 1. Connect the electrode arm to the base. 2. Connect the power cable to the connector cable to

Page 96 - 9 DATA MANAGEMENT

Instruction Manual CyberScan DO5000 87 At this point, the initial DO and all of the above information about the sample will be stored in the

Page 97

Instruction Manual CyberScan DO5000 88Measure - OURsetupTouch to begin measurementorstart Touch to access standardize modestd stdstartmg/

Page 98 - 10 DO FUNDAMENTALS

Instruction Manual CyberScan DO5000 891. Touch OUR on the main mode screen. 2. Touch setup to set parameters essential for the OUR calculati

Page 99 - 11 METER SPECIFICATIONS

Instruction Manual CyberScan DO5000 90Measure - SOURsetupTouch to begin measurementorstart Touch to access standardize modestd stdstartmg

Page 100

Instruction Manual CyberScan DO5000 911. Touch SOUR on the main mode screen. 2. Touch setup to set parameters essential for the SOUR calcula

Page 101

Instruction Manual CyberScan DO5000 927 CLEANING The touch screen should be kept as clean as possible to preserve optical properties. Attem

Page 102 - 12 ACCESSORIES

Instruction Manual CyberScan DO5000 93 9 DATA MANAGEMENT Your CyberScan DO 5000 meter is equipped with an RS-232 port for sending measuremen

Page 103 - 13 WARRANTY

Instruction Manual CyberScan DO5000 94To download data stored in memory to a printer, access the View Stored Data option in the setup screens

Page 104

Instruction Manual CyberScan DO5000 9510 DO FUNDAMENTALS Meter Operation The CyberScan DO 5000 meter uses the polarographic method of deter

Page 105

Instruction Manual CyberScan DO5000 96 11 METER SPECIFICATIONS Feature Details DO 5000 PCD 5500 Display 640X480 back-lit dot-matrix LCD  