Instruction Manual Cyanide Electrode
1
EUTECH INSTRUMENTS PTE LTD.
INSTRUCTION MANUAL
CYANIDE ION ELECTRODE
Page view 0
1 2 3 4 5 6 ... 17 18

Summary of Contents

Page 1 - INSTRUCTION MANUAL

Instruction Manual Cyanide Electrode 1EUTECH INSTRUMENTS PTE LTD. INSTRUCTION MANUAL CYANIDE ION ELECTRODE

Page 2 - TABLE OF CONTENTS

Cyanide Electrode Instruction Manual 10Using Table 3, the maximum ratio is: [Br-1] ———— = 5X103 [CN-1] [Br-1] = 5X103 [C

Page 3

Instruction Manual Cyanide Electrode 11 TABLE 4: Temperature vs. Values for the Electrode Slope Temp.(oC) "S" 0

Page 4 - TROUBLESHOOTING

Cyanide Electrode Instruction Manual 12Limits of Detection Cyanide levels from 5X10-6M to 1X10-2M cyanide can be measured with the cyanide electr

Page 5 - Sample Requirements

Instruction Manual Cyanide Electrode 13Electrode Operation Eutech Cyanide Ion Electrodes consist of a solid membrane containing a mixture of inor

Page 6

Cyanide Electrode Instruction Manual 14ions diffuse at different rates. When making measurements, it is important to remember that this potential

Page 7

Instruction Manual Cyanide Electrode 15- Measure correctly and accurately. - Review TROUBLESHOOTING HINTS. Standards and Reagents Whenever proble

Page 8

Cyanide Electrode Instruction Manual 16TROUBLESHOOTING HINTS Symptom Possible Causes Next Step Out of Range

Page 9 - Interferences

Instruction Manual Cyanide Electrode 17Low Slope or standards contaminated prepare fresh standards No Slope or

Page 10 - Temperature Influences

Cyanide Electrode Instruction Manual 18SPECIFICATIONS Concentration Range: 1X10-2M to 5X10-6M 260 to 0.1 ppm pH

Page 11 - Electrode Response

Cyanide Electrode Instruction Manual 2TABLE OF CONTENTS GENERAL INSTRUCTIONS ...

Page 12 - Electrode Storage

Instruction Manual Cyanide Electrode 3 EUTECH INSTRUMENTS PTE LTD. CYANIDE ION ELECTRODE INSTRUCTION MANUAL GENERAL INSTRUCTIONS Introduction E

Page 13 - Electrode Operation

Cyanide Electrode Instruction Manual 4 3. Cyanide Standard, 1X10-2M. To prepare this solution from your own laboratory stock, add 10 ml of ISA an

Page 14

Instruction Manual Cyanide Electrode 51. Prepare standard cyanide solutions whose concentrations vary by tenfold. Use either the 1.0X10-2M or the

Page 15

Cyanide Electrode Instruction Manual 6All samples must be aqueous and not contain organics which can dissolve the epoxy electrode body and/or the

Page 16

Instruction Manual Cyanide Electrode 7NaOH (ISA). After assuring that the meter is in the mV mode, lower the electrode tips into the solution. Aft

Page 17

Cyanide Electrode Instruction Manual 8 7. For low level measurements (below 8X10-6M or 0.2 ppm), place the rinsed, dried electrodes into a soluti

Page 18

Instruction Manual Cyanide Electrode 9 4. On a semi-logarithmic paper, plot the concentration (log axis) against the millivolt reading (linear ax